Fluke’s 9132 and 9133 infrared calibrators are designed to meet the precise needs of infrared thermometer and thermal imager users. These portable blackbody sources offer fast heating and cooling, stable temperatures, and radiometric accuracy—making them ideal for field and lab environments alike.
A blackbody calibrator (or blackbody source) is a device that emits thermal radiation with high emissivity , typically near 1.0. This makes it behave closely to an ideal blackbody, emitting radiation according to Planck’s law.
Blackbody calibrators are essential tools for ensuring the accuracy of:
This Model
Fluke 9132 Portable Infrared Calibrator. 50°C to 500°C
Key Features
- Accurate, reliable performance from –15 °C to 500 °C
- Small, compact design in a blackbody calibrator
- These blackbody calibrator units feature a temperature controlled measurement surface with a diameter of 2.25 in (57 mm).
- Intuitive, easy-to-read display that indicates when temperature is stable.
The Fluke Calibration 9132 Portable IR Calibrator, -30 to 150 °C, 115 V # 9132-156 is manufactured by Fluke Calibration. Valley Instrument Service is an authorized Fluke Calibration dealer offering full sales support, NIST-traceable calibration, and expert technical assistance.
Have questions about this product?
Our team can help with product selection, specs, pricing, and application support.
Contact Us or Call (661) 327-8681| Specifications | 9132; 9133 |
| Temperature range | 50°C to 500°C; (122°F to 932°F); –30°C to 150°C at 23°C ambient; (–22°F to 302°F at 73°F ambient) |
| Accuracy | ±0.5°C at 100°C; (±0.9°F at 212°F); ±0.8°C at 500°C; (±1.4°F at 932°F); ±0.4°C (±0.72°F) |
| Stability | ±0.1°C at 100°C; (±0.18°F at 212°F); ±0.3°C at 500°C; (±0.54°F at 932°F); ±0.1°C (±0.18°F) |
| Target size | 57 mm (2.25 in); 57 mm (2.25 in) |
| Target emissivity | 0.95 (±0.02 from 8 to 14 mm); 0.95 (±0.02 from 8 to 14 mm) |
| Resolution | 0.1°; 0.1° |
| Heating time | 30 minutes (50°C to 500°C); 15 minutes (25°C to 150°C) |
| Cooling time | 30 minutes (500°C to 100°C); 15 minutes (25°C to –20°C) |
| Computer interface | RS-232 I/O included |
| Power | 115 V AC (±10%), 3 A or; 230 V AC (±10%), 1.5 A,; switchable, 50/60 Hz, 340 W; 115 V AC (±10%), 1.5 A, or; 230 V AC (±10%), 1.0 A,; switchable, 50/60 Hz, 200 W |
| Size (H x W x D) | 102 x 152 x 178 mm; (4 x 6 x 7 in ); 152 x 286 x 267 mm; (6 x 11.25 x 10.5 in) |
| Weight | 1.8 kg (4 lb); 4.6 kg (10 lb) |
| NIST-traceable contact calibration | Data at 50°C, 100°C, 200°C, 250°C, 300°C, 400°C, and 500°C; Data at –30°C, 0°C, 25°C, 75°C, 100°C, 125°C, and 150°C |
Related Products

4180 Precision IR Calibrator, 152 mm (6 in) Target, -15 to 120 °C, 115 V
Precision Infrared Calibrator, 152 mm (6 in) target, -15 to 120 °C, 115 V

4181 Precision IR Calibrator, 152 mm (6 in) Target, 35 to 500 °C, 115 V
Precision Infrared Calibrator, 152 mm (6 in) target, 35 to 500 °C, 115 V

9133 Portable IR Calibrator, 50 to 500 °C, 115 V
Portable Infrared Calibrator with Blackbody Source, 50 to 500 °C, 115 V

1502A Tweener Thermometer Readout, 115 V
Thermometer, Tweener, 110V 50/60HZ