The E-DWT-H is used in the lab or instrument shop or can be taken into the field for performing in-situ calibrations and tests. This hydraulic pressure calibration system combines a high performance electronic reference pressure monitor and manually operated pressure generation and control hardware in a single, compact and rugged package. E-DWT-H offers the ease of use and precision of continuous, real time electronic pressure measurement with the simple and direct pressure control of high quality manual pressure hardware. E-DWT is named “electronic deadweight tester” because it is intended as a lightweight and easy to use, modern alternative to conventional piston-cylinder and weight based deadweight testers.
The E-DWT’s electronic reference pressure monitor is a special version of RPM4, designated RPM4-E-DWT. RPM4-E-DWT can be configured with one or two high precision quartz reference pressure transducers (Q-RPTs) with ranges from 7 MPa (1 000 psi) to 200 MPa (30 000 psi).
The RPM4-E-DWT is controlled locally by the operator using its front panel display, keypad and optional foot pedal or remotely by a computer using ASCII character command strings over its RS-232 interface.
Key Features
- Eliminates the need for weights and piston-cylinder changes.
- Real-time pressure indication without need for corrections.
- Set and read any pressure value exactly, operates in any unit of measure, and can switch easily between units.
- Offers ± 0.02% of reading total one-year measurement uncertainty.
The Fluke Calibration E-DWT Electronic Deadweight Tester, 140 MPa (20K PSI), Extended Range # 6532-140M is manufactured by Fluke Calibration. Valley Instrument Service is an authorized Fluke Calibration dealer offering full sales support, NIST-traceable calibration, and expert technical assistance.
Have questions about this product?
Our team can help with product selection, specs, pricing, and application support.
Contact Us or Call (661) 327-8681| To RPM4-E-DWT: | 12 V dc 1.2 A |
| To AC to DC power supply: | 100 V ac to 240 V ac, 50/60 Hz |
| Storage: | - 20 °C to 70 °C |
| Operating: | 10 °C to 40 °C |
| Storage: | 0 % to 100 % |
| Operating: | 0 % to 70 % |
| 1 Q-RPT: | 12 kg (26 lb) approximate |
| 2 Q-RPT: | 14 kg (30 lb) approximate |
| E-DWT footprint: | 41.4 cm W x 37.1 cm D (16.3 in x 14.6 in) |
| E-DWT height: | 26.9 cm (10.6 in), 33.6 cm (13.2 in) to max variable volume handle height |
| Pressure range | Up to 200 MPa (30k psi), depending on configuration |
| Operating medium | All 6531 and 6532 E-DWT Kits are shipped dry, standard preparation. Can be filled with di-ethyl-hexyl sebacate, silicon oils, propylene glycol, fully fluorinated liquids, partially fluorinated liquids, isopropyl alcohol and distilled water or mineral oil. |
| Reservoir capacity | 300 cc (18 in3) |
| Variable volume displacement | 3 cc (0.18 in3), 200 MPa (30,000 psi) maximum |
| Filling and priming pump displacement | 3.7 cc (0.23 in3) |
| TEST pressure connection | DH500 female, 200 MPa (30k psi) maximum working pressure. Adapters included for 1/8 in, 1/4 in, 3/8 in and 1/2 in NPT, 1/8 in, 1/4 in, 3/8 in and 1/2 in BSP, M20 x 1.5 and M14 x 1.5. Adapter maximum working pressure is 140 MPa (20k psi). Note: DH500 is a gland and collar type fitting for 6 mm (1/4 in) coned and left hand threaded tubes equivalent to AE F250C, HIP HF4, 9/16-18 UNF, etc. |
| Maximum Working Pressure: | Range of RPM4-E-DWT monitor’s Hi Q-RPT |
| Maximum priming pump pressure: | 700 kPa (100 psi) |
| Maximum Working Pressure: | with Lo Q-RPT selected: Range of RPM4-E-DWT monitor’s Lo Q-RPT |
| Communication ports: | RS-232 (COM1, COM2) |
| Warm up time | 15 minute temperature stabilization recommended from cold power up |
| Normal Operating Temperature Range | 10 °C to 40 °C (50 °F to 104 °F) |
| Resolution | Default: 0.01 % of active range; User adjustable to 1 ppm of Q-RPT maximum or 10 ppm of active AutoRange, whichever is larger |
| Precision1 | 200 MPa (30,000 psi) ranges: ± 0.018 % of reading or 0.0036 % of Q-RPT span, whichever is greater; ; All other ranges: ± 0.018 % of reading or 0.0018 % of Q-RPT span, whichever is greater |
| One year: | ± 0.0075 % of reading |
| Two year: | ± 0.015 % of reading |
| One year: | 200 MPa (30,000 psi) ranges: ± 0.02 % of reading or 0.004 % of Q-RPT span, whichever is greater; ; All other ranges: ± 0.02 % of reading or 0.002 % of Q-RPT span, whichever is greater |
| Two year: | 200 MPa (30,000 psi) ranges: ± 0.025 % of reading or 0.005 % of Q-RPT span, whichever is greater; ; All other ranges: ± 0.025 % of reading or 0.0025 % of Q-RPT span, whichever is greater |
This product is part of a family with 3 models. Select a model below to view details.
| Model | Price | |
|---|---|---|
6532-70M | Request Quote | View Details → |
6532-140MCurrent | Request Quote | Viewing |
6532-200M-B | Request Quote | View Details → |
Related Products

Adaptor Set, P3000 Standard
Adaptor Set, P3000 Standard
Conversion Weight Set, P3012/P3013 INH2O
Conversion Weight Set, P3012/P3013 INH2O
Conversion Weight Set, P3014 Bar
Conversion Weight Set, P3014 Bar
Conversion Weight Set, P3015 Bar
Conversion Weight Set, P3015 Bar