The 2271A features pressure ranges from -100 kPa to 20 MPa (-15 to 3000 psi), which covers the requirements of a wide range of gauges and sensors. Thanks to its modular design, the 2271A enables you to install two modules with different measurement ranges within the same chassis. You can purchase modules to match your current workload now; later, when your workload changes and grows, you can easily add ranges, swapping modules in and out according to your workload. This flexibility enables you to maintain your investment in the 2271A for years to come.
A built-in electrical measurement module (EMM) with HART capabilities enables you to perform closed loop, fully automated calibration on 4-20 mA devices such as smart transmitters, gauges and switches. Just set up the 2271A and then walk away to attend to other tasks.
The EMM supplies 24 V dc loop power for measuring mA and V dc. It has a built-in 250 Ohm resistor that can be toggled on or off, eliminating the need to have an external resistor to enable HART communications.
Key Features
- Everything you need to calibrate and test pressure transmitters and gauges from very low pressure to 3000 psi (20 MPa), all in one box
- Modular design supports multiple accuracy classes (0.02 % FS or 0.01 % Reading) with the ability to expand to cover wider pressure ranges
- Ideal for both beginner and expanding calibration laboratories.
- CPS Feature: Optional system to prevent contamination.
- Multilingual, intuitive interface and design.
The Fluke Calibration Pressure Calibrator, NPT, PM200-G2M/A100K, 2 MPa (300 PSI) # 2271A-N-G2M/A100K is manufactured by Fluke Calibration. Valley Instrument Service is an authorized Fluke Calibration dealer offering full sales support, NIST-traceable calibration, and expert technical assistance.
Have questions about this product?
Our team can help with product selection, specs, pricing, and application support.
Contact Us or Call (661) 327-8681| Power requirements | 0 V ac to 240 V ac, 47 Hz to 63 Hz |
| Fuse | T2A 250 V ac |
| Max power consumption | 100 W |
| Operating ambient temperature range | 15 °C to 35 °C |
| Storage temperature | -20 °C to 70 °C |
| Relative humidity | Operating: <80 % to 30 °C, <70 % to 35 °C |
| Vibration | MIL-T-28800D CLASS 3 |
| Altitude (Operation) | <2000 m |
| Ingress protection | IEC 60529: IP20 |
| Safety | IEC 61010-1, Installation Category II, Pollution degree 2 |
| Warmup time | 15 minutes typical |
| Weight (chassis only) | 15 kg (33.06 lbs) |
| Dimensions | Height: 2271A-NPT-HC20 305 mm (12 in) |
| Control Precision (Dynamic Mode) | PM200-BG2.5K; 0.005% Range Span |
| All other ranges | 0.001 % Range Span |
| Control turndown | 10:1 (Typical) |
| Low control point | 1 kPa (0.15 psi) absolute |
| Maximum particulate contamination | ≤ 1.25 micrometer (50 microinches) |
| Maximum moisture content | -50 °C dew point |
| Maximum hydrocarbon content | 30 ppm |
| Primary remote interfaces | Ethernet, RS232, USB |
| Connection | Standard 4 mm jack |
| Aux drivers | 4 external solenoid drivers |
| Model | Range (SI units); Range (Imperial units); Measurement mode; 1 year specification; (%FS) |
| PM200-BG2.5K | -2.5 kPa to 2.5 kPa; -10 inH₂0 to 10 inH₂0; gauge; 0.20 % |
| PM200-BG35K | -35 kPa to 35 kPa; -5 psi to 5 psi; gauge; 0.05 % |
| PM200-BG40K | -40 kPa to 40 kPa; -6 psi to 6 psi; gauge; 0.05 % |
| PM200-BG60K | -60 kPa to 60 kPa; -9 psi to 9 psi; gauge; 0.05 % |
| PM200-A100K | 2 kPa to 100 kPa; 0.3 psi to 15 psi; absolute; 0.10 % |
| PM200-BG100K | -100 kPa to 100 kPa; -15 psi to 15 psi; gauge; 0.02 % |
| PM200-A200K | 2 kPa to 200 kPa; 0.3 psi to 30 psi; absolute; 0.10 % |
| PM200-BG200K | -100 kPa to 200 kPa; -15 psi to 30 psi; gauge; 0.02 % |
| PM200-BG250K | -100 kPa to 250 kPa; -15 psi to 36 psi; gauge; 0.02 % |
| PM200-G400K | 0 kPa to 400 kPa; 0 psi to 60 psi; gauge; 0.02 % |
| PM200-G700K | 0 kPa to 700 kPa; 0 psi to 100 psi; gauge; 0.02 % |
| PM200-G1M | 0 MPa to 1 MPa; 0 psi to 150 psi; gauge; 0.02 % |
| PM200-G1.4M | 0 MPa to 1.4 MPa; 0 psi to 200 psi; gauge; 0.02 % |
| PM200-G2M | 0 MPa to 2 MPa; 0 psi to 300 psi; gauge; 0.02 % |
| PM200-G2.5M | 0 MPa to 2.5 MPa; 0 psi to 360 psi; gauge; 0.02 % |
| PM200-G3.5M | 0 MPa to 3.5 MPa; 0 psi to 500 psi; gauge; 0.02 % |
| PM200-G4M | 0 MPa to 4 MPa; 0 psi to 580 psi; gauge; 0.02 % |
| PM200-G7M | 0 MPa to 7 MPa; 0 psi to 1000 psi; gauge; 0.02 % |
| PM200-G10M | 0 MPa to 10 MPa; 0 psi to 1500 psi; gauge; 0.02 % |
| PM200-G14M | 0 MPa to 14 MPa; 0 psi to 2000 psi; gauge; 0.02 % |
| PM200-G20M | 0 MPa to 20 MPa; 0 to 3000 psi; gauge; 0.02 % |
This product is part of a family with 11 models. Select a model below to view details.
| Model | Price | |
|---|---|---|
2271A-N-G2M | Request Quote | View Details → |
2271A-N-G2M/BG200K | Request Quote | View Details → |
2271A-N-G2M/BG100K | Request Quote | View Details → |
2271A-N-G2M/A100KCurrent | Request Quote | Viewing |
2271A-N-G7M | Request Quote | View Details → |
2271A-N-G7M/G700K | Request Quote | View Details → |
2271A-N-G20M | Request Quote | View Details → |
2271A-N-G20M/G2M | Request Quote | View Details → |
2271A-N-G20M/G250K | Request Quote | View Details → |
2271A-N-BG200/2.5K | Request Quote | View Details → |
Related Products

Lines and Fittings Kit for 2271A-NPT-HC20
Lines and Fittings Kit for 2271A-NPT-HC20

Portable Gas Pressure Calibrator, Line Cords US/EU/UK/China/Australia
Portable Gas Pressure Calibrator, Line Cords US/EU/UK/China/Australia
Reusable Transit Case for 227X
Reusable Transit Case for 227X

2700G Reference Pressure Gauge, -12 to 100 psi (-80 to 700 kPa)
Reference Pressure Gauge, -12 to 100 psi (-80 to 700 kPa)